Common faults of vacuum hot press furnace and their elimination methods

2022-10-18 09:28:02
times

The common faults in the use of vacuum hot press furnace and their elimination methods

High vacuum can not be pumped but does not leak (refers to the vacuum meter for a long time value is not moving)

Possible causes:

1. High vacuum regulation aging

2, diffusion pump aging

Solution: Replace the vacuum regulator; Replace the diffusion pump oil.


Low vacuum can not be pumped but does not leak (refers to the vacuum meter for a long time value is not moving)

Possible causes:

1. The working current of low vacuum regulator is incorrect.

2, mechanical pump aging

Solution: Adjust the working current of vacuum regulator; Replace the mechanical pump oil.


Thermocouple automatic entry and exit failure, alarm lamp alarm

Possible causes:

1. Failure of inner and outer limits.

2, in and out of the motor does not rotate.

Solution: Adjust or replace the internal and external limit device; Test whether the motor is broken, such as bad replacement of the same type of motor. (Note) Before heating the electric furnace, observe whether the internal limit indicator light is on and should be on.


The temperature will not rise after a period of use

Possible cause: The resistance value of the heater increases.

Solution: According to the instrument instructions to increase the output power of the instrument; Severe aging can only be replaced with a new heater


The overcurrent indicator on the panel of the control cabinet lights up and gives an alarm

Possible cause: The heater has a short circuit

Solution: Open the furnace cover to observe whether there is a short circuit heater, troubleshooting.


Overtemperature alarm light and alarm

Possible causes:

1, the furnace temperature is over temperature

2. Instrument setting is wrong

Solution: Reset the overtemperature alarm parameters according to the instructions


Water cut-off indicator light and alarm

Possible causes:

1. The water valve is not opened

2. The water pressure is too low

Solution: Open the water valve; Turn up the water pressure.

Article content from the network, if you have any questions please contact me to delete!


Vacuum hot press furnace

Tags

Recently Viewed:

Related products

news.related_news

Shenyang Weitai Technology Development Co., Ltd

Tel:024-78992751

Fax:024-78992761 

Address of factory:East Tieling County, Tieling City, Liaoning Province Nw Intersection of 1st Street and Xidi Road (New High Line)


Copyright © Shenyang Weitai Technology Development Co., LTD record number: Mainly engaged inVacuum hot press furnace,Vacuum hot pressure diffusion welding furnace,Ceramic hot press furnace, Welcome to inquire!